RTD Meter Specifications
RTD Meters
Specifications
& Accessories
6
(800) 421-2853 FAX (828) 658-0728 www.palmerwahl.com
392 SERIES METER ACCESSORIES
DA-4 Silicone Paste MA-150 Battery Recharger, 120V AC 50/60Hz
DA-6 Belt-Clip Meter Case with Hand Strap MA-150E Battery Recharger, 220V AC 50/60Hz, European
DA-10 Nylon Safety Wrist Strap 11681-1 Spare NiCad Battery Pack
12423-06 Shock-Proof Attache-Style Instrument Case 11681-2
Discontinued
12232 Standard 9V Alkaline Battery
METER OPTIONS
392NIST
Certificate with 3 Standard Calibration
Points between 32°F (0°C) & 500°F (250°C)
Meter calibration includes one probe.
392NIST-1
Certificate with 3 User-specified Calibration
Points between -40°F (-40°C) & 600°F (315°C)
Meter calibration includes one probe.
08/13
Calibration Services Available
170 Years of Continued Innovation
392A SERIES RTD METER SPECIFICATIONS
Model
392AHP
3
92AM
392AMX
3
92AM-S
392AMX-S
392AVX
R
ange
-
290
° t
o 1450°F
-180° to 788°C
-
60
° t
o 752°F
-51° to 400°C
-
60
° t
o 752°F
-51° to 400°C
Digital Display
Bright 0.33"
LED
392AM
Bright 0.33" high LED
Discontinued
Red, 0.4" high LCD
3
92AMX
Red, 0.4" high LCD
Resolution 0.1° below 375°F (190°C). Auto Ranges to above 375°F (190°C).
Meter Accuracy at 77°F ±0.1°F/C ±1 digit at 32°F (0°C)
System Accuracy
(Meter with 202
immersion probe)
±0.1°F ±1 digit at ice point,
±0.25% reading thereafter
±0.3°F at ice point,
±0.5% reading thereafter
Repeatability ±0.2°F
Peak Hold Microcomputer calculated, switch to “P” to activate
Ambient Operating Range
0° to 130°F (-18° to +55°C)
Temperature Coefficient 0.01 deg/deg over ambient range
FM (Factory Mutual)
Approved*
Discontinued Discontinued Discontinued
Not
Available
Power
11681-1
6.25 V
NiCad Battery
11681-1
6.25 V
NiCad Battery
11681-2
Discontinued
9V Alkaline battery
(NEDA 1604A)
Battery Life 6 hours. per charge
392AM
8 hours per charge
Discontinued
8 hours.
per charge
392AMX/392
10 hours per charge
Low Battery Indicator Yes
Noise Rejection
Greater than 40 dB at 60 Hz increasing at 20 dB per decade. Spurious noise
attenuated by internal 16 Hz filter. 392AHP includes special noise rejection
program for use in presence of rampant EMF such as spark ignition engines.
Size 3" W x 6" L x 1.5" D (7.5 cm x 15 cm x 3.75cm)
Weight 12 oz. (340 grams) 12 oz. (340 grams) 7 oz. (198 grams)
All 392A Models Available in °F, °C, or Dual Scale, add F”, C” or Dto Model Number.
See our New TM612 RTD Thermometer, and
TM630 Thermocouple and RTD Thermometer featured on page 3.
WARRANTY