TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Corning M103 pH Meter
MFG. MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20SC–42
Page 2 of 2
4.2 PH Indication Tests
4.2.2 TI control(s):
MODE switch to pH
TEMP C to 25 CALIB.
for TI display of 7.00 ck ( ) NA
4.2.3 (DC millivolt source)
(mV) (pH) (pH)
295.000 2.00 1.97 to 2.03
177.000 4.00 3.97 to 4.03
59.000 6.00 5.97 to 6.03
–59.000 8.00 7.97 to 8.03
–177.000 10.00 9.97 to 10.03
–295.000 12.00 11.97 to 12.03
4.3 PH Calibration Tests
4.3.4 TI control(s):
MODE sw to PH
Adjust TEMP to std.
digital thermometer indication ck ( ) NA adj. to std. Ind.
Adjust CALIB for TI display
of 7.00 ck ( ) NA adjust
4.3.6 Adjust TEMP to std.
digital thermometer indication ck ( ) NA adj. to std. Ind.
Adjust 2% for TI display
of 4.01 ck ( ) NA adjust
4.3.8 pH measurement (pH) (pH)
4.01 3.98 to 4.04
4.3.10 pH measurement (pH) (pH)
10.01 9.98 to 10.04
NAVAIR 17–20SC–42
10