
TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Beckman Instruments 2000 Ion Analyzer
MFG. MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20SC–42
Page 3 of 3
4.3.6 TI conrol(s):
Set MID SCALE VALUE
control to 4
Release STBY pushbutton
Adjust SLOPE
control until needle is at
pH 4, center scale ck( ) NA
4.3.8 pH Measurement (pH) (pH)
4.01 3.4 to 4.08
4.3.10 pH Measurement
Set MIDSCALE VALUE
control to 7 ck( ) NA
(pH) (pH)
10.01 9.4 to 10.08
NAVAIR 17–20SC–42
17
Comentarios a estos manuales