2500/1370 Series Type K
Thermocouple Meters
Thermocouple Meters
2500/1370 Series
10
(800) 421-2853 • FAX (828) 658-0728 • www.palmerwahl.com
05/08
Calibration Ser
vices A
vailable
170 Y
ears of Continued Innovation
2
500 Meter
R
ange -100° to 2500°F
1370 Meter Range -70° to 1370°C
• Meter options for °F or °C range, display type and power
• Meter accuracy is ±0.2% of reading ±1 digit
• Maxi-Temp
®
automatically holds peak reading
• Accepts all type K probes with ANSI mini-connectors
• Microprocessor circuitry for thermocouple linearization
Heat-Prober
®
Thermocouple Thermometers and Probes are
rugged and reliable instruments made for process and
maintenance temperature measurements.
The pocket-sized microprocessor-based Heat-Prober Type K
thermometer updates readings three times per second with an
accuracy of ±0.2% of reading ±1 digit in 1°F or °C resolution.
The precision cold junction is compensated with a unique,
permanently calibrated circuit.
The Wahl Maxi-Temp peak hold switch, available on all models,
permits the Heat-Prober to hold and display the highest
temperature reached. This function is ideal where the operator
must concentrate on the probe when making the measurement
and where an instantaneous transient peak is to be recorded.
Maxi-Temp meters also can be switched to normal operation.
Advanced electronics are housed in ABS plastic, thoroughly
tested for environmental integrity and subjected to 4-foot drop
tests, battery life and running life tests. All units are calibrated
with simulated EMF inputs and powered during a high-
temperature burn-in cycle.
See Meter specifications on page 12.
See page 12-13 for 2500/1370 Series meter,
pr
obe, and accessor
y money-saving kits.
See pages 14-17 for interchangeable thermocouple probes.
WARRANTY
2500MX Type K Thermocouple Heat-Prober
Model 2500M (-100° to 2500°F) features a bright .33" high red
LED display for normal to low lighting conditions. It operates on
a rechargeable 6.25V NiCad battery pack. Includes battery
recharger/AC line adapter.
Model 1370M as above with -70° to 1370°C range.
Model TCMR-K °F/°C switchable.
Model 2500MX (-100° to 2500°F) features a .5" high red LCD
display for normal to bright lighting conditions. It operates on a
rechargeable 6.25V NiCad battery pack. Includes battery
recharger/AC line adapter.
Model
1370MX
as above with -70
°
to 1370
°C range.
Model TCMR-KX °F/°C switchable.
Model 2500MVX (-100° to 2500°F) features a .5" high red LCD
display for normal to bright lighting conditions and is power
ed
by an easily replaceable single standard 9V battery.
Model 1370MVX as above with -70° to 1370°C range.
Model TCMR-KVX °
F/°C switchable.
NIST TRACEABLE THERMOCOUPLE CALIBRATORS
This thermocouple meter
accessory allows
calibration over the
meter’s full range.
Useful for quick field
diagnostics, it assures
meter accuracy whenever
a measur
ement is in
question.
Accuracy is ±0.3°F/°C or
0.2% of reading,
whichever is greater.
User sends the
calibrator
only back to the factory
for NIST Certification.
One Calibrator maintains
NIST traceability for all
in-house meters.
THERMOCOUPLE CALIBRA
TOR
MODELS
Model T
ype
Calibration Points
T
A30AFK
K 40
°
F - 140
°
F - 170°
F
T
A50AFK
K 100
°F - 155°F - 200°F - 250°F - 300°F
TA70AFK K 130°F - 650°F - 1300°F - 1900°F
TA70ACK K 70°C - 400°C - 800°C - 1200°C
TA70AFT T 32°F - 80°F - 150°F - 250°F
TA-70AFK Precision Calibrator in use
with 2500MX Thermocouple Meter
NIST Certification included