RTD Meter Specifications
RTD Meters
Specifications
& Accessories
6
(800) 421-2853 • FAX (828) 658-0728 • www.palmerwahl.com
WARRANTY
392 SERIES METER ACCESSORIES
DA-4 Silicone Paste MA-150 Batter
y Recharger, 120V AC 50/60Hz
DA-6 Belt-Clip Meter Case with Hand Strap MA-150E Battery Recharger, 220V AC 50/60Hz, European
DA-10 Nylon Safety Wrist Strap 11681-1 Spare NiCad Battery Pack
DA-60 Shock-Proof Attache-Style Instrument Case 11681-2 Spare NiCad Pack for -S Models
12232 Standard 9V Alkaline Battery
METER OPTIONS
392NIST
Certificate with 3 Standard Calibration
Points between 32
°
F (0
°C) & 500°F (250°C)
Meter calibration includes one probe.
392NIST
-1
Certificate with 3 User-specified Calibration
Points between -40
°
F (-40
°C) & 600°F (315°C)
Meter calibration includes one probe.
05/08
Calibration Ser
vices A
vailable
170 Y
ears of Continued Innovation
392 SERIES RTD METER SPECIFICATIONS
M
odel
392HP
392M
3
92MX
392M-S
3
92MX-S
392VX
Range
-290° to 1450°F
-180
° to 788°C
-60° to 752°F
-51
° to 400°C
-60° to 752°F
-51
° to 400°C
Digital Display
Bright 0.33"
LED
392M/392M-S
Bright 0.33" high LED
Red, 0.4" high LCD
3
92MX/392MX-S
Red, 0.4" high LCD
Resolution 0.1° below 375°F (190°C). Auto Ranges to 1˚ above 375°F (190°C).
Meter Accuracy at 77°F ±0.1°F/C ±1 digit at 32°F (0°C)
System Accuracy
(Meter with 202
immersion probe)
±0.1°F ±1 digit at ice point,
±0.25% reading thereafter
±0.3°F at ice point,
±0.5% reading thereafter
Repeatability ±0.2°F
Peak Hold Microcomputer calculated, switch to “P” to activate
Ambient Operating Range
0° to 130°F (-18° to +55°C)
Temperature Coefficient 0.01 deg/deg over ambient range
FM (Factory Mutual)
Approved*
Class I & II,
Div. 1, Groups:
D, F, & G
Class I & II,
Div. 1, Groups:
D, F, & G
Class I & II,
Div. 1, Groups:
C, D, E, F, & G
Not
Available
Power
11681-1
6.25 V
NiCad Battery
11681-1
6.25 V
NiCad Battery
11681-2
6.25 V
NiCad Battery
9V Alkaline battery
(NEDA 1604A)
Battery Life 4.5 hours. per charge
392M/392M-S
10 hours per charge
50 hours.
per charge
392MX/392MX-S
50 hours per charge
Low Battery Indicator Yes
Noise Rejection
Greater than 40 dB at 60 Hz increasing at 20 dB per decade. Spurious noise
attenuated by internal 16 Hz filter.
392HP includes special noise rejection
program for use in presence of rampant EMF such as spark ignition engines.
Size 3" W x 6" L x 1.5" D (7.5 cm x 15 cm x 3.75cm)
Weight 12 oz. (340grams) 12 oz. (340grams) 7 oz. (198grams)
* FM Approval Groups: C (Ethylene Gas), D (Methane, Propane), E (Metal Dust), F (Coal Dust), and G (Grain Dust)
All 392 Models A
vailable in
°
F
, °C, or Dual Scale, add “
F”, “C” or “D” to Model Number
. Dual Scale not available on -S Models.
See our New TM612 RTD Thermometer, and
TM630 Thermocouple and RTD Thermometer featured on page 3.
Vista de pagina 5
1 2 3 4 5 6 7 8 9 10 11 ... 23 24

Comentarios a estos manuales

Sin comentarios